User manual
Table Of Contents

Operator's Manual 
8  WA1K2K-OM-E RevB 
Fuse Replacement 
(For Only WaveAce 2000 Models Only) 
Set the instrument Standby (power) switch to Standby mode, flip the 
mains power switch to the OFF (0) position, and disconnect the power 
cord before inspecting or replacing the fuse. Open the black fuse holder 
(located at the rear of the instrument below the main power switch) 
using a small, flat-bladed screwdriver. Remove the old fuse, replace it 
with a new 5 X 20 mm T rated 1.25 A/250 V fuse, and reinstall the fuse 
holder. 
WARNING 
For continued fire protection at all line voltages, replace fuse with the 
specified type and rating only. Always disconnect the power cord before 
replacing the fuse. 
Calibration 
The recommended calibration interval is one year. Calibration should be 
performed by qualified personnel only. 
Cleaning 
Clean only the exterior of the instrument, using a damp, soft cloth. Do 
not use chemicals or abrasive elements. Under no circumstances allow 
moisture to penetrate the instrument. 
Avoid electrical shock hazard by unplugging the power cord from the AC 
outlet before cleaning. 
 WARNING - Electrical Shock Hazard 
  No operator serviceable parts inside. 
  Do not remove covers. 
  Refer servicing to qualified personnel. 










