User manual
WaveAce Remote Control
22
920836 Rev A
ACQUISITION – PEAK DETECT, PDET
Description
The PEAK_DETECT command switches ON or OFF the peak detector built into the acquisition
system.
The PEAK_DETECT? query returns the current status of the peak detector.
Command Syntax
Peak_DETect <state>
<state> : = {ON, OFF}
Query Syntax
Peak_DETect?
Response Format
PDET <state>
Example
The following instruction turns on the peak detector:
PDET ON
ACQUISITION - STOP
Description
The STOP command immediately stops the acquisition of a signal. If the trigger mode is AUTO
or NORM, STOP places the oscilloscope in STOPPED trigger mode to prevent further acquisition.
Command Syntax
STOP
Related Commands
ARM_ACQUISITION, TRIG_MODE, WAIT, FORCE_TRIGGER