User manual

ArbStudio
72
922244-00 Rev A
The same marker operations are available by right clicking on one of the
markers in the list and selecting the desired function.
DDS (CARRIER EDITOR)
In DDS mode the only available marker identifies a single amplitude value
on the carrier waveform. During the waveform generation, when the
instrument output becomes higher than the marker value, a
leading/trailing edge on the Trigger Out signal is generated.
The Trigger Out signal remains high/low while the output signal remains
higher than the marker amplitude, at which point it switches.
The leading/trailing edge for the Trigger Out signal can be selected on the
Channel Settings Trigger Out tab.
If the selected instrument is configured as the Master, a marker can also
send a synchronous signal on an Expansion Bus line.