48
922244-00 Rev A
Trigger OUT Math Tab
Enabled channels can have a logical combination set between all trigger
out events.
The possible operations include Force Low (force to low logic level signal),
Force High (force to high logic level signal), AND, OR, XOR, NAND, NOR,
and XNOR.
Results are made visible on the BNC Trigger OUT Connector and on the
Trigger Out signal of the digital connector (Pod A/B).
If the Pattern Generator option is available, click the Toggle button to
switch the BNC Trigger Out selection from the left pair to the right (or vice
versa).