User manual
Operator’s Manual
923064-00 Rev A 7
Electrical Ratings
Maximum Input Voltage 1000 V rms (1400 V DC + peak AC)
Installation (Overvoltage) Category III
Pollution Degree 2
Measurement Category III (CAT III) is applicable for measurements performed in the building installation.
Pollution Degree 2 is applicable to an operating environment where normally only dry non-conductive pollution
occurs. Occasionally a temporary conductivity caused by condensation must be expected.
General Characteristics
Operating Temperature -10º C to 40º C
Storage Temperature -30º C to 70º C
Altitude, Operating up to 2000 m (6560 ft)
Cable Length 1.3 m
Weight (probe only) 42 g
Power requirement Four internal 1.5 V AA size batteries or
6 V d.c. / 60 mA mains adaptor
(Not supplied )
Dimensions 6.6" (168 mm) x 2.4" (62 mm) x 0.79" (20 mm)
(excluding casing )
Cables BNC, 95 cm ( RG58 / U ), Input Lines: 45 cm
(PVC, double insulation )
Accessories 2 x Safety Hooks, 4 mm compatible (1 red, 1 black)
Weight 9.35 oz ( 285 g ) excluding batteries and casing
Safety Certification EN61010-1:2010 ; EN61010-031/A1:2008
1000V CAT III Approval