User Manual

MultiClamp-Meter Pro
17
GB
Red
Black
When measuring resistance, to avoid the risk of
the results of a measurement being distorted, there
must be no traces of dirt, oil, solder spray or other
contamination on the test prods.
!
When measuring resistance, the components must be
zero-potential.
!
Resistances can only be measured
correctly in isolation; therefore,
the components might need to be
disconnected from the remainder
of the circuit.
!
Resistance measurement8
DCA Zero
7
1. Connect the red test prod to the V Ć input socket (12) and
the black test prod to the COM input socket (8).
2. Turn the rotary switch to position „Ć“.
3. Connect the measuring contacts to the object to be tested.
4. The measured value is shown on the LC display. If the
measured value is < 30 Ć, a signal sounds.
2. 3.1.
Black
Red
When measuring voltage, the resistance measurement
and continuity test functions must not be activated.
!
4. The measured value is shown on the LC display. By pressing
the „MAX/MIN“ button, the minimum and maximum values
are measured and displayed.
When measuring DC voltages, reset the value by pressing
the „MAX/MIN“ button and holding it down for 2 seconds
to calibrate the electronic measuring equipment. By pressing
the „MAX/MIN“ button repeatedly, the minimum and
maximum values are measured displayed.