Operating instructions

21
GB
CoatingTest-Master
Switch the device to calibration mode as described in step 9 and keep the
"ESC" key pressed until the display shows the measuring mode. The display
may show the following messages regarding calibration:
Take the following steps to perform a zero
calibration:
1. Turn on the measuring device without allowing the measuring head to
contact any metal objects.
2. Activate "individual measurement" mode (item 4 Options)
3. Place the measuring head vertically onto the provided uncoated base
pattern (always calibrate on clean uncoated surfaces)
4. Put down the measuring device after measuring.
5. Keep the "zero" key pressed for 2 seconds
6. Repeat steps 3 to 5 several times.
7. Zero calibration is complete. Disable the calibration mode again.
Zero calibration
10
cal
no one-point or two-point calibration available
cal zero
cal 1~2
one-point or two-point calibration available
zero
no zero calibration available
zero Y
zero calibration available
The measuring device determines the average of the last 5 zero calibra-
tions and overwrites the oldest value. Zero calibrations are recommen-
ded before each new measurement.
!