Manual

ThermoCamera HighSense (Pro)
1. Set emissivity coefcient to 1.0
2. Blur the object out of focus (Refer to Section 15.1)
3. Aim the camera in the opposite direction
of the actual measured object
4. Determine the average temperature
5. Set the average temperature as the reection
temperature
Infrared measurements of specic objects can be affected by reection from other objects nearby or even
from the ambient air because the measured object cannot be fully isolated. The reection temperature
setting can help compensate for the interference. The reection temperature is generally the same as the
ambient temperature. However, if larger objects with signicantly different temperatures (approx. > 20 °C)
are in close proximity to the measured area, its effects must be taken into account. For this purpose,
follow these steps:
High humidity can lead to the lens of the thermal
imaging camera misting over which means the infrared
radiation is not fully received. The infrared radiation is
absorbed on the way to the lens by the atmosphere
and distributed by the water vapour suspended in the
air, amongst other things (relative humidity).
These inuences should be taken into consideration,
particularly when the distance to the measured object
exceeds around 30 m. This can have a negative
inuence on the measuring accuracy. The parameter
can be set from 10% to 100% to compensate for
the inuence of the relative humidity. Thick fog can
also affect the measurement as the water droplets
in the transmission path allow less infrared radiation
through.
Relative humidity
Reection Temperature
7.2
Parameter: Relative humidity
7.1
Parameter: Reection temperature
33
EN