Corp Computer Accessories User Manual

LaserSpeed 8000-3 Instruction Handbook
RS-232/RS-422 Communication
Part No. 93463 / Drawing No. 0921-01561 Page 159 of 221 Revision A (Sep 2007)
79,+000000741,000012344,15,63,3125<CR>
79,+000000864,000012342,15,63,3125<CR>
These readings indicate the length increasing from +0.370 to 0.864, with a velocity near
+12.345, a Quality Factor of 15, Status of 63, and an internal gauge temperature of 31.25ÂșC.
TV Read One Line of TT Data
TV<CR> Data Output Command
Response Format: Variable (Depends on TT Value)
This command will cause the LS8000-3 to transmit one line of data in the format configured
by the TT command. See the TT command for details.
. Stop Real-time Output
.<CR> Command
This command stops the LS8000-3 from sending real-time measurement data. This
command has the same effect as just sending a <CR>.