Datasheet
EL-USB-TC
K, J & T-Type Thermocouple Data Logger
Issue 16 10/2018 Page 3 of 3
www.lascarelectronics.com/data-loggers
EL-WIN-USB
Lascar’s EasyLog control soware is available as a free download from www.easylogusb.com. Easy to install and use, the control
soware is compable with 32-bit and 64-bit versions of Windows 7, 8 & 10. The soware is used to set up the logger, download,
graph and annotate data or export in Excel, PDF and jpeg formats.
The soware allows the following parameters to be congured:
▪ Logger name
▪ Measurement parameter (°C or °F)
▪ Logging rate (user selectable between 1 second and 12 hours)
▪ High and low alarms
▪ Immediate and delayed logging start
Download the latest version of the soware free of charge from www.easylogusb.com
BATTERY INFORMATION
Replacement
We recommend that you replace the baery annually, or prior to logging crical data. Only use 3.6V ½AA lithium metal baeries.
The data logger does not lose its stored readings when the baery is discharged or replaced; however, the data logging process will
stop and will not resume unl the baery is replaced and the logger restarted by EL-WIN-USB or an EL-DataPad.
Before replacing the baery, remove the data logger from the PC. Please note that leaving the data logger plugged into the USB
port for extended periods will cause some of the baery capacity to be lost.
Passivaon
If le unused for extended periods of me lithium metal baeries, including those used in the EasyLog range of data loggers,
naturally form a non-conducve internal layer prevenng them from self-discharge and eecvely increasing their shelf life. When
rst installed in the data logger, this may cause a momentary drop in the baery voltage (the Transient Minimum Voltage) as
the internal layer is broken down, resulng in the data logger reseng. Inserng the baeries in the data logger and leaving it
connected to a PC for about 30 seconds will remove this layer. Aer this, remove and re-install the baeries to reset the data
logger. Overall baery life will not be aected.
WARNING