Datasheet

EL-USB-TC
K, J & T-Type Thermocouple Data Logger
Issue 16 10/2018 Page 3 of 3
www.lascarelectronics.com/data-loggers
EL-WIN-USB
Lascars EasyLog control soware is available as a free download from www.easylogusb.com. Easy to install and use, the control
soware is compable with 32-bit and 64-bit versions of Windows 7, 8 & 10. The soware is used to set up the logger, download,
graph and annotate data or export in Excel, PDF and jpeg formats.
The soware allows the following parameters to be congured:
▪ Logger name
▪ Measurement parameter (°C or °F)
▪ Logging rate (user selectable between 1 second and 12 hours)
▪ High and low alarms
▪ Immediate and delayed logging start
Download the latest version of the soware free of charge from www.easylogusb.com
BATTERY INFORMATION
Replacement
We recommend that you replace the baery annually, or prior to logging crical data. Only use 3.6V ½AA lithium metal baeries.
The data logger does not lose its stored readings when the baery is discharged or replaced; however, the data logging process will
stop and will not resume unl the baery is replaced and the logger restarted by EL-WIN-USB or an EL-DataPad.
Before replacing the baery, remove the data logger from the PC. Please note that leaving the data logger plugged into the USB
port for extended periods will cause some of the baery capacity to be lost.
Passivaon
If le unused for extended periods of me lithium metal baeries, including those used in the EasyLog range of data loggers,
naturally form a non-conducve internal layer prevenng them from self-discharge and eecvely increasing their shelf life. When
rst installed in the data logger, this may cause a momentary drop in the baery voltage (the Transient Minimum Voltage) as
the internal layer is broken down, resulng in the data logger reseng. Inserng the baeries in the data logger and leaving it
connected to a PC for about 30 seconds will remove this layer. Aer this, remove and re-install the baeries to reset the data
logger. Overall baery life will not be aected.
WARNING
Handle lithium metal baeries carefully, observe warnings on baery casing. Dispose of in accordance with local regulaons.