Datasheet

EL-USB-TC
K, J & T-Type Thermocouple Data Logger
Issue 16 10/2018 Page 1 of 3
www.lascarelectronics.com/data-loggers
This standalone data logger measures and stores more than 32,000 temperature readings from a J, K or T-type thermocouple at a
resoluon of 0.5°C (1°F). It comes supplied with a K-type thermocouple capable of measuring from 0 to +200°C (+32 to +392°F).
Your applicaon will determine which probe is most suitable based on temperature range, accuracy, form and price. A wide variety
of alternave thermocouples are available. Please call Lascar for vendor recommendaons.
The user can easily set up the logger and view downloaded data by plugging the data logger into a PCs USB port and using the free
EasyLog soware. Data can then be graphed, printed and exported to other applicaons for detailed analysis.
The data logger is supplied with a lithium metal baery which typically gives 2 years’ logging life.
Compable with K, J and T type thermocouples with
miniature thermocouple plug connecon
Stores over 32,000 readings
EasyLog soware available as a free download
Supplied with 1.5m K-type thermocouple probe with
0 to 200°C (32 to 392°F) measurement range
Logging rates between 1 second and 12 hours
Immediate and delayed logging start
User-programmable alarm thresholds
Status indicaon via red and green LEDs
Probe measurement range 0 to 200°C (32 to 392°F) K-type (supplied)
-200 to +1350°C (-328 to +2462°F) K-type
-200 to +1190°C (-328 to +2174°F) J-type
-200 to +390°C (-328 to +734°F) T-type
Internal resoluon 0.5°C (1°F)
Accuracy (overall error) ±1°C (±2°F) (data logger only - thermocouple
error not included)
Logging rate User selectable between 1 second & 12 hours
Operang temperature range -10 to +40°C (+14 to +104°F) (data logger only)
Baery Life 2 years (at 25°C and 1 minute logging rate)
Readings 32,510
Dimensions 118 x 27 x 27mm (4.64 x 1.06 x 1.06”)
SPECIFICATIONS ACCESSORIES
INCLUDED IN THE BOX
Replacement baery
Baery
Replacement K-type
thermocouple probe
K-type thermocouple
probe
Handheld datalogger
programmer & collector
Mounng Bracket
BAT 3V6 1/2AA
BAT 3V6 1/2AA
EL-P-TC-K
EL-P-TC-K
EL-DataPad
EL-USB WALL BRACKET
CALIBRATION CERTIFICATES NOW AVAILABLE
Lascar now oers a Traceable Calibraon Cercate Service on Temperature Data Loggers. Using reference equipment
which has been calibrated by a UKAS/NIST accredited laboratory and using apparatus traceable to naonal or
internaonal standards. For more informaon, please see www.lascarelectronics.com.

Summary of content (3 pages)