LED STATUS INDICATION
The EL-USB-1 features two LEDs: a green/
red LED to indicate logging and alarm
status, and an amber LED to indicate a
possible problem.
In normal operaon the green LED will
ash, but will change to red if an alarm
condion has been triggered. Using
EL-WIN-USB it is possible to set the alarm
to remain acve even if the reading has
returned to normal, in which case the
alarm LED will connue to ash red.
This ‘Hold’ feature in the soware
ensures the user is noed that at some
point an alarm level has been exceeded,
without needing to download the data.
BATTERY INFORMATION
Replacement
We recommend that you replace the baery annually, or prior to logging crical data. Only use 3.6V ½AA lithium metal baeries.
The data logger does not lose its stored readings when the baery is discharged or replaced; however, the data logging process will
stop and will not resume unl the baery is replaced and the logger restarted by EL-WIN-USB or an EL-DataPad.
Before replacing the baery, remove the data logger from the PC. Please note that leaving the data logger plugged into the USB
port for extended periods will cause some of the baery capacity to be lost.
Passivaon
If le unused for extended periods of me lithium metal baeries, including those used in the EasyLog range of data loggers,
naturally form a non-conducve internal layer prevenng them from self-discharge and eecvely increasing their shelf life. When
rst installed in the data logger, this may cause a momentary drop in the baery voltage (the Transient Minimum Voltage) as
the internal layer is broken down, resulng in the data logger reseng. Inserng the baeries in the data logger and leaving it
connected to a PC for about 30 seconds will remove this layer. Aer this, remove and re-install the baeries to reset the data
logger. Overall baery life will not be aected.
WARNING
Handle lithium metal baeries carefully, observe warnings on baery casing. Dispose of in accordance with local regulaons.
Green double ash
The data logger is not currently logging, but is primed to start at a later date
and me (delayed start)
Green single ash
The data logger is currently logging. No alarm
Green and Amber single ash
The data logger is full and has stopped logging. No alarm
Red single ash
The data logger is currently logging. Low alarm
Red and Amber single ash
The data logger is full and has stopped logging. Low alarm
Red double ash
The data logger is currently logging. High alarm
Red and Amber double ash
The data logger is full and has stopped logging. High alarm
Amber single ash
The baery is running low. Data logging will connue unl the baery voltage
drops below 2.8V
No LEDs ash
The data logger is stopped, the baery is empty or there is no baery
MEASUREMENT ACCURACY
± 0.0
± 0.5
± 1.0
± 1.5
± 2.0
-40 -200 20 40 60 80
± 0.0
± 0.5
± 1.0
± 1.5
-350 50 80
Temperature (°C)
Accuracy (°C)
± 0
± 2
± 4
± 6
± 8
± 10
0102030405060708090 100
Relative Humidity (%RH)
RH (%RH)
Maximal RH-accuracy at 25°C per sensor type
± 0.0
± 0.5
± 1.0
± 1.5
± 2.0
± 2.5
± 3.0
-40-20 020406080
Temperature (°C)
T (°C)
Maximal T-accuracy per sensor type
EL-USB-2-LCD