Data Sheet
www.lascarelectronics.com/data-loggers www.lascarelectronics.com/data-loggers
EL-USB-2
Temperature, Humidity and Dew Point
Data Logger
Issue 15 01/2020 Page 1 of 4
This standalone data logger measures and stores over 16,000 temperature and humidity readings from -35 to +80°C (-31 to +176°F)
and 0 to 100%RH range at a resoluon of 0.5°C (1°F) and 0.5%RH.
The user can easily set up the logger and view downloaded data by plugging the data logger into a PC’s USB port and using the
free EasyLog soware. Data, including calculated dew point, can then be graphed, printed and exported to other applicaons for
detailed analysis.
The data logger is supplied with a lithium metal baery, giving up to three years’ logging life. The logger is protected against ingress
from water and dust to IP67 standard when the cap is ed.
▪ -35 to +80°C (-31 to +176°F) and 0 to 100%RH measurement range
▪ Stores over 16,000 readings for both temperature and humidity
▪ EasyLog soware available as a free download
▪ Logging rates between 10 seconds and 12 hours
▪ Immediate and delayed logging start
▪ User-programmable alarm thresholds for both temperature and humidity
▪ Status indicaon via red/green LEDs
▪ Environmental protecon to IP67
Temperature Measurement range -35°C to 80°C (-31°F to 176°F)
Internal resoluon 0.5°C (1°F)
Accuracy (overall error)* 0.45°C (1.04°F) typical (5 to 60°C)
Long term stability <0.02°C (0.04°F) / year
Relave Humidity Measurement range 0 to 100%RH
Internal resoluon 0.5%RH
Accuracy (overall error)* 3%RH (20 to 80%RH)
Long term stability <0.25%RH / year
Dew Point Accuracy (overall error)* 1.7°C typical
(-35 to 80°C, 40 to 100%RH)
Logging rate User selectable between 10
seconds & 12 hours
Operang temperature range -35 to +80°C (-31 to +176°F)
Baery life 3 years (at 25°C and 1 minute
logging rate)
Readings 16,382 temperature, 16,382
relave humidity
Dimensions 108 x 25 x 22mm
(4.25 x 0.98 x 0.86”)
* The overall error takes in to account the sensor accuracy (as shown on page 3) and the resoluon of the
data logger
SPECIFICATIONS ACCESSORIES
Replacement baery
Handheld data logger
programmer & collector
BAT 3V6 1/2AA
EL-DataPad
INCLUDED IN THE BOX
Baery
Mounng Bracket
BAT 3V6 1/2AA
EL-USB WALL BRACKET
CALIBRATION CERTIFICATES NOW AVAILABLE
Lascar now oers a Traceable Calibraon Cercate Service on Temperature Data Loggers. Using reference equipment
which has been calibrated by a UKAS/NIST accredited laboratory and using apparatus traceable to naonal or
internaonal standards. For more informaon, please see www.lascarelectronics.com.