Use Instructions
Table Of Contents
46
Technical Data
en
Accuracy
Flow ±10 % (for max. flow)
±50 mmHg (for maximum pump pressure)
n/a
Pressure
Deficit
Dimensions Width x Height x Depth 210 mm x 148 mm x 260 mm
Weight 3.7 kg
Interfaces/ports
IN/OUT signal for components Service interface (D-Sub connector 9-pin, only for service, not ESD
protected.)
Mains connection IEC-60320-1 C14
RFID Transponder Technology Transmit/Receive Frequency Range: 13.56 MHz ± 0.424 MHz
Transceiver class: Class I
RF Output Power: -10.83 dBμA/m at 10 m/ 32.8 ft
Type of Antenna: Inductive Loop Antenna
Antenna loop area: 0.00032 m²
Modulation: Amplitude-shift keying (ASK)
Mode of Operation (Simplex / Duplex): Duplex
Essential performance Providing a irrigation pressure to generate a flushing flow. The
maximum irrigation pressure is 450 mmHg +-10 %.
Providing a suction pressure to generate fluid suction. The maxi-