11 Technical Data
Type or model designation: PP110
Manufacturer information: W.O.M. WORLD OF MEDICINE GmbH
Salzufer 8, 10587 Berlin
Mains voltage range 100-240 V~
Supply frequency range 50/60 Hz
Fuse designation 2 x T 3.15 AH, 250 V
Power consumption Current (A) Voltage (V) Power consumption
(VA)
Upper voltage range
Normal operation 0.31 240 73
Peak 0.33 240 79
Lower voltage range
Normal operation 0.61 100 58
Peak 0.65 100 65
Protection class (I, II, III) I
Application part type (B, BF, CF) BF
Defibrillator protection (yes/no) No
Protection class (IP code) IP41
Classification (I, IIa, IIb, III) acc. to Appendix IX of the
Europ. MDD
IIa
Conformity with the following standards:
(in their current versions)
IEC 60601-1, EN 60601-1, ANSI/AAMI ES 60601-1, CAN/
CSA C22.2 No. 60601-1
Operating conditions 10 to 40 °C/50 to 104 °F
30 to 75 % relative humidity
70 to 106 kPa air pressure
3000 m maximum altitude above sea level for device use
Possible use with explosive anesthetic gases This device is not designed for use with flammable anesthetic
gases (Class AP) or flammable anesthetic gases with oxygen (Class
APG).
Storage conditions -20 to +60 °C/-4 to +140 °F
10 to 85 % relative humidity
70 to 106 kPa air pressure
Transportation conditions -20 to +60 °C/-4 to +140 °F
10 to 85 % relative humidity
70 to 106 kPa air pressure
Maximum sound level < 80 dB(a)
Maximum negative suction pressure -60 kPa
Maximum pump pressure 450 mmHg
Maximum irrigation performance 2.0 l/min (withT0505-01/T0506-01/T0500-01 and T0502-01)
3.0 l/min (withT0508-01)
Precision (repeatability)
Flow ±10 % (for maximum flow)
±50 mmHg (for maximum pump pressure)
n/a
Pressure
Deficit