Use Instructions
42
Technical Data
en
11 Technical Data
Type or model designation: PP120
Manufacturer information: W.O.M. WORLD OF MEDICINE GmbH
Salzufer 8, 10587 Berlin
Mains voltage range AC 100-240 V ~
Supply frequency range 50/60 Hz
Fuse designation 2 x T 3.15 A, H, 250 V~ , UL-recognized
Power consumption: Current (A) Power consumption (VA)
Normal operation 100V AC/60 Hz 0.72 72
Normal operation 240V AC/50 Hz 0.38 93
Max. Power consumption - 95
Protection class (I, II, III) I
Application part of the type (B, BF, CF) BF
Defibrillator protection (yes/no) No
Protection class (IP-Code) IP21
Classification (I, IIa, IIb, III) acc. to Appendix IX of the
Europ. MDD .
IIa
Conformity with the following standards:
(in the currently valid version)
EN 60601-1 / IEC 60601-1
EN 60601-1-2 / IEC 60601-1-2
Operating conditions 10 to 40 °C / 50 to 104 °F
30 to 75 % rel. humidity
70 to 106 kPa air pressure
3000 m max. altitude above sea level for device use
Possible use with explosive anesthetic gases This device is not designed for use with flammable anesthetic gases
(Class AP) or flammable anesthetic gases with oxygen (Class APG).
Storage and transportation conditions -25 to +55 °C / -13 to +131 °F
10 to 85 % rel. humidity
70 to 106 kPa air pressure
Max. Sound level < 80 dB(a)
Maximum negative suction pressure -60 kPa
Max. pump pressure 450 mmHg
Maximum irrigation performance 2.0 l/min (with T0505-01/T0506-01/T0500-01 and T0502-01)
3.0 l/min (with T0508-01)
Accuracy
flow ±10 % (for max. flow)
±50 mmHg (for maximum pump pressure)
n/a
Pressure
Deficit
Accuracy
flow ±10 % (for max. flow)
±50 mmHg (for maximum pump pressure)
n/a
Pressure
Deficit
Dimensions Width x Height x Depth 380 mm x 148 mm x 388 mm
Weight 6.0 kg
Interfaces/ports/connections