Use Instructions

Annual Inspection
41
en
10.2 Irrigation Test
Preparation
1. Switch device on using the power switch.
2. Insert the tube set.
3. Connect the irrigation tube with both irrigation fluid bags. The fluid bags
have to be suspended 1.5 m above the device.
4. Start the irrigation process:
Press the irrigation start/stop key (LED lights up).
5. Open the flushing valve to fill the tube set completely with liquid.
6. Close the flushing valve.
Testing
1. Start test
Empty measuring cup.
2. Hold tube into measuring cup, open outlet, and stop at 30 seconds.
3. End test:
Press the irrigation start/stop key (LED goes out). The measuring cup must
contain approx. 1.5 l (±10 %) of water.
10.3 Suction Test
1. Connect the suction system including suction container and instrument.
Close stopcock if using instrument with stopcock.
2. Press the suction start/stop key (LED lights up). The device generates a nega-
tive pressure inside of the suction system.
The vacuum pump stops when a negative pressure of max. -60 kPa (-0.60 bar)
has been reached.
The suction system has a leak, if the vacuum pump is not deactivated. Check
the suction system for leaks and repeat test.
3. Starting suction:
Open the instrument.
The vacuum pump automatically reactivates, if the min. negative pressure is
exceeded.
4. End suction:
Press the suction start/stop key again (LED goes out).
Each successfully conducted test must documented with the test log.