Use Instructions
36
Care and Maintenance
en
9.5 Cleaning the Device
1. Switch device off.
2. Disconnect device from power supply. For this, pull the mains plug from the
mains socket.
The concentration of the used disinfectant depends on the information provided
by the manufacturer of the disinfectant. Wipe the surface of the device with a
soft cloth moistened with the disinfectant. Make sure moisture does not enter
the device.
The manufacturer recommends Meliseptol® rapid as disinfectant.
WARNING!
Cleaning the device
Do not sterilize the device.
9.6 Care of the Reusable Tube Set (NOT FOR SALE IN USA)
The device can be used with a reusable tube set that is listed as an accessory (see
chapter 12 Accessory List) Please comply with the following information when
using a reusable tube set.
9.6.1 Cleaning the Reusable Tube Set (NOT FOR SALE IN USA)
WARNING!
The reusable tube set (NOT FOR SALE IN USA) is made of silicone, polysulfone
(PSU), and stainless steel. For cleaning and disinfection, use only pH-neutral or
slightly alkaline cleaning agents (e.g. neodisher MediClean 2.0 %), disinfectants
(e.g. Lysetol V 8 %), and drying and rinsing agents approved for the tube set ma-
terials.
Using unsuitable agents (e.g. neodisher MediKlar rinsing agent) may damage
the tube system and especially the PSU connectors.
Reprocessing restrictions The manufacturer has tested and approved the reusable tube set (NOT FOR SALE
IN USA) for a specific number of reprocessing cycles. Always comply with the in-
Fig. 9-1 Opening the fuse holder