Use Instructions
Using the Device during Surgery
33
en
8 Using the Device during Surgery
WARNING!
Functional test
The functional test must be performed prior to each surgery.
WARNING!
Select an irrigation fluid suitable for the subsequent medical procedure.
Before surgery1. Tube set and instrument are filled with irrigation fluid. The suction system is
connected. The stopcocks are closed.
2. Press the suction start/stop key (LED lights up).
The suction system is pre-evacuated.
Only then can the suction procedure be started with the instrument.
During Surgery1. Press the irrigation start/stop key (LED lights up).
2. Insert instrument into abdomen.
Start the irrigation process:
Actuate the trumpet valve or open the stopcock.
3. Starting the suction process:
Actuate the trumpet valve or open the stopcock.
After surgery
NOTE!
The manufacturer recommends emptying the tube set using the pump when em-
ploying saccharine irrigation fluids (HF use). Remove irrigation fluid bags and in-
strument first.
1. Press the irrigation start/stop key and the suction start/stop key (LED goes
out).
2. Use the power switch to turn device off.
3. Remove the tube sets.
NOTE!
Comply with hygiene rules and regulations when disposing of the tube set, col-
lected fluid, and the suction container.