5.11 Replacing the Suction Container
CAUTION!
Full suction containers must be replaced immediately without stopping surgery.
1. Stop the suction (see chapter 5.11 Replacing the Suction Container).
Press the suction start/stop key (LED goes out).
2. Actuate the trumpet valve or Open the stopcock to vent the system.
3. Remove tubes from filled suction container.
4. Reattach tubes to empty suction container.
Close stopcock if using instrument with stopcock.
5. Press the suction start/stop key (LED lights up).
6. Starting suction:
Actuate the trumpet valve or open the stopcock.
5.12 Removing a Vacuum Tube
1. Remove the drain tube (4) from the instrument (5) (see Fig. 5-14 Suction
system with suction-/Irrigation instrument or Fig. 5-15 Suction system with
standard instrument).
2. Remove the suction container (3) from the drain tube (4).
3. Remove the vacuum tube (1) from the suction container (3).
4. Remove the hygiene filter from the vacuum tube (1).
5. Remove the vacuum tube from the device.
5.13 Switching Device Off
Ending operation of the pump safely:
1. Press the ON/OFF switch (see Fig. 5-1 "Device front" (11))
2. Disconnect device from power supply. For this, pull the mains plug from the
mains socket (see Fig. 5-2 "Device rear" (17)).