Use Instructions
24
Operating the Device
en
The day tube The day tube includes the connectors to the fluid bags and to the hygiene con-
nector (Fig. 5-10 "Patient tube elements" see (12)) of the patient tube.
The day tube may be used the entire day of surgery. However, it is designed for a
maximum of 10 uses.
CAUTION!
The day tube must be replaced at the end of the day of surgery but not later than
after 24 hrs.
A new sterile cap (included with the patient tube in the delivery) must be screwed
onto the hygiene connector (Fig. 5-9 (10)) of the day set immediately after the
surgical procedure to protect against contamination. This sterile protector re-
mains connected to the day tube until the next surgery.
The patient tube The patient tube has a connector on one side to connect with the day tube (hy-
giene connector (12)). Luer lock connector (14) to connect to an instrument such
as a trocar, for example, is fitted to the other side. . Connect the hygiene connec-
tor (12) on the patient tube with the matching piece on the day tube (Fig. 5-9
(10)).
The patient tube set has to be replaced for every patient to prevent cross-con-
tamination. The day tube remains connected to the device after each surgery.
The used patient tube must be disposed of after each surgery.
Fig. 5-9 Day tube set elements
(1) Protective caps
(2) Tap spikes
(3) Tube clamps
(4) Y-connector
(5) Irrigation tube
(6) Ring
(7) Roller tube
(8) Pressure chamber with mem-
brane and transponder
(9) Instrument tube
(10) Hygiene connector
(11) Protective cap
(2)
(6)
(8)
(4)
(10)
(9)
(7)
(1)
(3)
(5)
(11)
Fig. 5-10 Patient tube elements
(11) Protective cap for hygiene connec-
tor
(12) Hygiene connector
(13) PVC tube, 285 cm
(14) Luer lock connector with screw cap
(15) Protective cap for Luer lock
(15)
(14)(13)
(12)
(11)