Use Instructions

Initial Device Startup
15
en
CAUTION!
ME Device in Rack
The ME device may not be used when stacked or stored directly adjacent to or
with other devices. If it should be necessary to operate the device close to
stacked with other devices, the ME device or ME system should be monitored to
ensure it works properly as configured.
CAUTION!
Equipment should be positioned such that power cord can be easily disconnect-
ed.
CAUTION!
Ventilation of the device
Avoid device overheating. Ensure free air circulation especially to the bottom
and rear of the device (rear panel distance of at least 10 cm).
CAUTION!
Before switching on the device, sufficient time must have passed to adjust to the
indoor climate.
CAUTION!
Position the device in such a way that it is easy to operate and switch off.
CAUTION!
Place the device outside the sterile field.
CAUTION!
Position of the user
To avoid a malfunction, the user must be positioned correctly
within a display viewing angle of ±50° to operate the device
up to 2 m/6.5 ft from the device front for monitoring the actual values
Mains connection
CAUTION!
Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause errors
and malfunctions and may destroy the device.
Make sure the connection data and technical specifications of the power supply
comply with DIN VDE or national requirements. The mains connection cable may
be plugged only into a properly installed, grounded safety wall socket (shock-
proof socket) (see DIN VDE 0100-710). Read the device label located in rear of de-
vice (type plate) to determine the operating voltage of the device.
Grounding contactThe power connection must be equipped with a grounding contact. Use the orig-
inal power cable (if included in scope of delivery) to establish a connection be-
tween the mains wall socket and the non-heating device plug located in the rear
of the device.