Use Instructions

General Information
11
en
3.3.2 Precautions
CAUTION!
Electrical Interference
(See chapter 13 "Electromagnetic Compatibility"). Electrical interference with
other devices or instruments was practically eliminated when developing this
devices and none was detected during testing. However, if you still detect or sus-
pect such interference, please follow these suggestions:
Move this, the other or both devices to a different location
Increase distance between used devices
Consult an electro-medical expert
CAUTION!
Endoscope
The device may only be connected with endoscopes designed for and featuring
the technical specification permitting such a combined use. Any utilized endo-
scopes must comply with the most recent versions of IEC 60601-2-18 and
ISO 8600. Combining/connecting with other devices generates a medical electri-
cal system (MES). The system configurator is responsible for compliance with the
standard IEC 60601-1 / EN 60601-1 in its latest version.
CAUTION!
Mains voltage
Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause errors
and malfunctions and may destroy the device.
CAUTION!
Indoor climate
Before switching on the device, sufficient time must have passed to adjust to the
indoor climate.
CAUTION!
Power cables
Any power cables employed by the user that are not provided by the manufac-
turer must meet the safety requirements of the national standards in the respec-
tive current valid version.
CAUTION!
Ventilation of the device
Avoid device overheating. Ensure free air circulation especially to the bottom
and rear of the device (rear panel distance of at least 10 cm).
CAUTION!
The pressure chamber diaphragm is fragile and can be replaced if damaged. Re-
placement diaphragms are supplied with reusable tubes. The reusable tube sets
must be prepared before each use. See Chapter 9.6 Care of the Reusable Tube Set
(NOT FOR SALE IN USA) for further instructions.