Use Instructions

10
General Information
en
WARNING!
Make sure the tubes for the suction-irrigation instrument do not have any kinks
when installed. Make sure the tubes are not crimped or clamped off.
WARNING!
Filter
The vacuum tube with integrated filter is designed for max. 30 days. Replace the
vacuum tube if it is obviously contaminated. The filter prevents that body fluids
enter the interior of the device. Please note that the filter may reduce the suction
capacity.
WARNING!
Do not touch
Avoid touching the roller wheel. Risk of injury!
WARNING!
Serum sodium concentration
It is also necessary to monitor the concentration of sodium in the blood of the pa-
tient to prevent electrolyte disturbances. Monitoring of the concentration of so-
dium in the blood must be performed by the physician and is not performed or
supported by the system.
WARNING!
Air Embolisms
An air embolism can be the result of air contained in the tube set or connected
instrument reaching the patient. Make sure that there is always fluid in the bag
to prevent air from being aspirated into the patient.
WARNING!
Position the device in such a way that it is easy to operate and switch off.
WARNING!
An authorized service technician must check the device if the specified parame-
ters and tolerances are exceeded.