Use Instructions

General Information
9
en
WARNING!
The indicated drying time depends on several variables, including the following:
Altitude, humidity, type of packaging, preconditioning, size of chamber, mass of
load, and placement in chamber. Users must verify that the drying time set in
their autoclave yields dry surgical equipment when using the method of saturat-
ed steam sterilization described here.
WARNING!
Remove the membrane before cleaning.
WARNING!
Make sure to connect tubes as indicated by the matching connectors.
WARNING!
The day tube remains connected to the device after each surgery. The patient
tube must be disposed of immediately after surgery. A new sterile cap (included
with the patient tube) must be screwed onto the hygiene connector at the day
tube after each surgery. This sterile protector remains connected to the day tube
until the next surgery.
WARNING!
The device is only intended for use with flexible fluid containers. Do not use glass
containers as they might break. Fluid cannot flow quickly enough due to the vac-
uum being generated inside of the bottle. Risk of implosion.
WARNING!
Select an irrigation fluid suitable for the subsequent medical procedure.
WARNING!
Do not touch patient and the ON/OFF switch at the same time.
WARNING!
ME System (Medical Electrical System)
Use only parts and/or devices from ME systems (see Chapter 15 Glossary) in pa-
tient environments in compliance with the standard IEC60601-1 in the respec-
tive currently valid version.
WARNING!
ME System (Medical Electrical System)
Connect only items that have been specified as part of the ME system or speci-
fied as being compatible with ME system.