Use Instructions
8
General Information
en
WARNING!
Contamination
Do not use device or vacuum tube if signs of contamination are detected. Fluid
that has penetrated can be detected, for example, by dirt encrustations on the
connections for the suction or at the exit of the vacuum pump. Make sure the de-
vice or/and accessories can no longer be operated until a qualified service tech-
nician conducts the appropriate tests and repairs.
WARNING!
Different default settings of the warning message for identical or similar devices
in the operating room may cause a risk due to conflicting acoustic signals.
WARNING!
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
WARNING!
Reprocessing of sterile disposable products
Infection hazard for patients and/or users and impairment of product function-
ality due to reuse. Risk of injury, illness or death due to contamination and/or
impaired functionality of the product! Do not reprocess the product.
WARNING!
The vacuum tube sets for this device contain diethylhexylphthalate (DEHP),
which is classified as toxic to reproduction according to the EU Directive 1272/
2008/EEC on Classification, Labeling and Packaging of Dangerous Substances.
DEHP may impair fertility and may cause harm to the unborn child. Therefore,
this product must not be used for unauthorized applications. When applied
within the intended use, the potential risk to pregnant or breastfeeding women
as well as to children resulting from the DEHP contained in this product is not
critical.
WARNING!
The reusable tube set (NOT FOR SALE IN USA) is made of silicone, polysulfone
(PSU), and stainless steel. For cleaning and disinfection, use only pH-neutral or
slightly alkaline cleaning agents (e.g. neodisher MediClean 2.0 %), disinfectants
(e.g. Lysetol V 8 %), and drying and rinsing agents approved for the tube set ma-
terials.
Using unsuitable agents (e.g. neodisher MediKlar rinsing agent) may damage
the tube system and especially the PSU connectors.
WARNING!
Please check the reusable tube (NOT FOR SALE IN USA) for signs of damage after
sterilization and before use. Never use a tube showing signs of damage, especial-
ly brittleness and perforation.