Charging the Wireless Unit
Charge the battery before using the unit for the first time and when the LOW BATT LED
illuminates steady red, indicating that approximately 10% of battery capacity remains.
You can continue using the wireless unit for data transfer while it is charging.
1 Plug the charging adapter into the charger receptacle on the front of the
wireless unit.
CAUTION: Use only the charger provided by LifeScan.
2 Plug the other end of the charging adapter into an AC power outlet.
CAUTION: Do not crimp the charging adapter cable—either leave the tote lid open
during charging or remove the wireless unit from the tote.
As the unit fast charges, the CHARGING LED blinks. Fast charging is complete
when the LED illuminates steady green. You may unplug the adapter from the
wireless unit at this point. Or, to completely charge the unit, allow the charging to
continue for an extended period—up to a maximum of 15 hours is recommended.
Charging the wireless unit longer than 15 hours may shorten the battery life.
3 Unplug the charging adapter from the wireless unit.
Cleaning the Wireless Unit
Clean the wireless unit if dirt or blood is present, or as defined by your institution’s
infection control policies.
1 Clean the exterior of the unit using a cloth moistened with a 10% bleach
solution. Follow with a cloth moistened with water to remove residual bleach.
NOTE: For a complete list of approved cleaning agents, refer to the Specifications
section of the SureStepFlexx Meter Operator’s Guide.
2 Dry the unit thoroughly.
CAUTION: Do not get water inside the wireless unit. Never immerse the unit or hold it
under running water.
ethernet.book Page 19 Friday, December 2, 2005 4:28 PM