Use Instructions

PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 84 of 101
Table 40 Custom Sterile Sleeve Wired Configuration Instructions
8.
While holding the HFC USB Hatch, turn the
small sterile bag inside-out to safeguard the
HFC USB Hatch while not in use.
9.
Connect the HFC USB Cable
to the HFC.
Position the HFC USB Cable
90° Type-C connector such
that it points towards the
inside of the HFC.
10.
Link the HFC USB Cable to the ECH Link Cable, while
observing sterile technique.
Connect ECH Link Cable to ECH PC USB Port per Section
6.3.2.3 Step 8.
Before proceeding to the next step in this instruction,
verify successful connection between the HFC and ECH
per Section 6.3.2.3 Steps 8 through 14.