5.5.2. Sterilizing Protective Storage Cradle
Do not continue with sterilizing the Protective Storage Cradle unless the cleaning process described in
section 5.5.1 is completed. Effective sterilization can only be achieved after completing the cleaning
process.
Steam sterilize the Protective Storage Cradle according to the following instructions.
Table 26 Protective Storage Cradle Sterilization Instructions
1.
Place Protective Storage Cradle in a double-wrapped sterilization configuration
2.
Subject PSC to full cycles of either
1) Four (4) minutes of exposure time at 132°C using prevacuum air removal
2) Three (3) minutes of exposure time at 134°C using prevacuum air removal
3.
Subject PSC to dry times of either
1) 20 to 30 minutes of dry time after 132°C with four (4) minutes exposure using prevacuum air
removal following Four (4) minutes of exposure time at 132°C using prevacuum air removal
2) 20 minutes of dry time after 134°C with three (3) minutes exposure using prevacuum air
removal following three (3) minutes of exposure time at 134°C using prevacuum air removal
4.
The PSC is now disinfected and ready for use. Observe sterile technique when transporting
equipment from re-processing locations to points of use.