Use Instructions
PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 57 of 101
Table 22 Dark Imaging Box Disinfection Instructions
2.
Thoroughly wipe crevices and areas that are difficult to reach. Where required, fold or bunch
Intermediate Level Disinfection Wipes to push into crevices, corners, and areas that are difficult to
reach.
3.
Use additional Intermediate Level Disinfection Wipes as necessary to ensure ALL SIDES and surfaces
of the DIB including its Base, Cradle, Adjustment Wheel, Ruler, and Imaging Platform have been wet.
4.
Once wet, allow the DIB to dry.
5.
The DIB is now disinfected and ready for use. Observe sterile technique when transporting
equipment from re-processing locations to points of use.
5.4. CCD – Contact Charging Device Cleaning and Disinfection
Warning
Do not clean or disinfect the Contact Charging Device, its power supply, or its
cables when energized. Unplug all cable connections before cleaning or
disinfecting. Cleaning or disinfecting when energized may be hazardous to the
operator and/or destructive to the device(s).
Caution
Prior to disinfection, ensure surfaces are cleaned to remove organic matter as
described in the sections below.
Thoroughly clean and disinfect all surfaces of the Contact Charging Device, the CCD Power Supply, and the
cables using Intermediate Level Disinfection Wipes.
5.4.1. Cleaning CCD
Table 23 Contact Charging Device Cleaning Instructions
1.
Wear impervious gloves such as latex gloves when handling Intermediate Level Disinfection Wipes.
2.
Before cleaning, ensure the CCD is UNPLUGGED from the CCD Power Supply, and ensure the CCD
Power Supply is UNPLUGGED from the electrical outlet.
3.
Use at least two new Intermediate Level Disinfection Wipes to clean and remove all visible soil and
debris from the CCD, the CCD Power Supply, and all cables.
4.
Thoroughly clean crevices and areas that are difficult to reach. Where required, fold or bunch
Intermediate Level Disinfection Wipes to push into crevices, corners, and areas that are difficult to
reach.
5.
As necessary, use additional Intermediate Level Disinfection Wipes if they become soiled during the
cleaning process.
6.
In a well-lit area, visually inspect (at a distance of approximately 12 inches) the CCD, the CCD Power
Supply, and all cables to ensure all surfaces are clean. If necessary, repeat the cleaning steps above
until the CCD, the CCD Power Supply, and all cables are clean.