Use Instructions
PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 35 of 101
Warning Description
The Handheld Fluorescence Camera utilizes auto-exposure to optimize the exposure settings
based on lighting conditions. Images captured on the HFC are relative, not quantitative, and
may not represent absolute fluorescence levels.
Avoid strong electrical or electromagnetic fields, e.g. mobile phones, microwave ovens, etc.;
which may result in temporary imaging error or inaccuracy.
3.3.2. General Cautions
Caution Description
Reflective objects (e.g. retractors or other metal objects) can produce red light reflections that
can appear similar to PpIX Fluorescence.
Sources of ambient light reflecting off of surfaces in the field of view may contaminate HFC
images.
Always verify ambient lighting conditions using the Ambient Light Sensor before performing
Fluorescence Imaging.
Always respect the Range Finder icons when performing Imaging.
Do not operate the Handheld Fluorescence Camera using eyewear with color tint, polarization,
or other factors which may distort the interpretation of colors on acquired images.
The Handheld Fluorescence Camera should not be used by individuals confirmed to be color
blind as the typical green and red color information comprising the acquired images may be
misinterpreted.
Ensure the optical head is clean prior to imaging. Acquired images may appear distorted, out
of focus, or otherwise compromised if the optical head is not clean prior to imaging.
Prior to charging the device, ensure the CFC Power Cable is undamaged before plugging into a
wall outlet.
Only use cables and power supplies provided by SBI ALApharma Canada Inc. with the Eagle
V1.2 Imaging System.
The optical head may get warm after prolonged use. The device will disable activated LEDs
(white light or blue-violet light) if device temperature exceeds 41°C or 106°F.