User's Manual
DD399 MolecuLight DX® User Manual for Wi-Fi Certification
Page: 32 of 42
Inactive
Active
2. Tap the Video/Still toggle to set it to Video capture mode.
3. Aim the device parallel to the wound.
4. Ensure the distance from the wound is acceptable and that the range finder indicator is green.
5. Centre the target in the middle of the field of view.
6. Tap the screen to focus on the target.
7. Tap START RECORDING. The device will begin recording and display a counter on the field of view to
count down from the maximum allowable video length of 30 seconds.
8. (Optional) Tap STOP RECORDING before the 30 seconds to capture a shorter video.
9. A thumbnail will be displayed on the bottom left of the Imaging screen when the video is saved.
10. (Optional) Tap the thumbnail to review your video in full screen. To exit full screen mode, tap Close.
11. Continue to capture videos or tap Back to return to the Wound screen.
12.5 Fluorescence Images and Videos
Fluorescence digital imaging capabilities allow you to acquire (in real-time) and document intrinsic fluorescence
emission of dermatological wounds between 500-545 nm and 600-665 nm when illuminated by violet 405 nm
excitation light under acceptable environmental lighting conditions.
A dark environment is required to capture a good quality image in Fluorescence mode, so turn off room lights.
The software allows you to capture a still image or a video in Fluorescence mode. Still images are captures in *.jpeg
format and videos are captured in *.mov format.
To capture a Fluorescence still image:
1. Turn off the room lights.
2. From the Imaging screen in Standard mode, tap the FL available icon to enter Fluorescence mode.
FL
available
Inactive
Active
3. Tap the Video/Still toggle to set it to Still capture mode.
4. Aim the device parallel to the wound.
5. Ensure the distance from the wound is acceptable and that the range finder indicator is green.
6. Centre the target in the middle of the field of view.
7. Tap the screen to focus on the target.