Use Instructions
PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 99 of 101
7. Workflow
7.1. Unpack
1) Unpack and remove all Eagle V1.2 Imaging System components.
2) Charge HFC with CCD per Section 6.3.5.
3) Set up the ECH per Section 6.3.2.1 and verify HFC connection.
7.2. Imaging Workflow
1) Clean, disinfect, and sterilize all Eagle V1.2 Imaging System components per Section 5.
2) Mount the CSS on the HFC per Section 6.3.1
3) Verify successful HFC connection with ECH per Section 6.3.2.2 or 6.3.2.3.
4) Perform imaging per Section 6.2.2.2
5) Use the Dark Imaging Sheet and Dark Imaging Box as necessary per Section 6.3.3 and Section
6.3.4.
7.3. Post Imaging
1) Dispose of CSS, DIS, and DIB Covers.
2) Clean, disinfect, and sterilize Eagle V1.2 Imaging System components per Section 5.
8. Care and Maintenance
8.1. Transport and Handling
Per section 6.3.6 Store the Handheld Fluorescent Camera in the Protective Storage Cradle when not in
use.
9. Performance Specifications
9.1. HFC – Handheld Fluorescence Camera
• Operating Temperature: 18 – 25 °C
• Operating Relative humidity: 10 – 70%
• Operating Atmospheric pressure: 89 – 102 kPa
• Height: 13 cm
• Width: 19 cm
• Depth: 13 cm
• Weight: 591 g
• Input (Charging Current, Voltage): 2.5 A, 5 VDC
• Battery Capacity: 6500 mAh
• Operating Voltage: 3.7 VDC