Use Instructions
PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 94 of 101
Table 43 External Communications Hub Wired Instructions
7.
Before proceeding, ensure the Custom Sterile Sleeve (CSS) has been configured for ECH Wired
Configurations per Section 6.3.1.2, up until Step 10.
8.
Connect ECH Link Cable to the ECH PC USB Port.
When a successful connection is made, ECH notifies “First Device is
connected to the Hub: wired”.
9.
ECH notifies “Hub is ready for Wireless setup – Connect Second
Device”.
If not connecting two (2) HFCs, select “Enter” and move to the next
step.
If connecting two (2) HFCs, connect second ECH Link Cable into the
ECH PC USB Port.
10.
“Second Device is connected to the Hub: wired”.
11.
ECH notifies “Hub Setup Complete”.
12.
Review the HFC Display Screen.
Connected Handheld Fluorescent Camera(s) will show “ECH:
Connected USB” in the top left corner.
13.
Press “Enable Casting” to cast the HFC Display Screen onto the ECH
Display Monitor Wirelessly.
14.
Verify the ECH Display Monitor is casting the HFC Display Screen.
6.3.3. DIS – Dark Imaging Sheet
The DIS provides a black non-reflective, non-fluorescent standardized surface background imaging in the
sterile field or with the Dark Imaging Box per Section 6.3.4.
Table 44 Dark Imaging Sheet Instructions
1.
When used in the sterile field, place the DIS on a flat surface with the
side labelled “This Side Down / Ce Côté En Bas” facing down.