6.3. Accessories Overview and Operation
6.3.1. Custom Sterile Sleeve (CSS) Instructions for Use
The Custom Sterile Sleeve can be sheathed around the Handheld Fluorescence Camera in two (2) ways to
accommodate both ECH Wireless (Section 6.3.2.2) and ECH Wired (Section 6.3.2.3) Configurations.
Observe sterile technique when mounting the CSS as described in the next Sections 6.3.1.1 and 6.3.1.2.
6.3.1.1. Custom Sterile Sleeve Wireless Configuration
Table 39 Custom Sterile Sleeve Wireless Configuration Instructions
1.
Ensure the External Communications Hub (ECH) is set up for ECH Wireless Configurations per Section
6.3.2.1.
2.
Pull the CSS out of its
packaging.
Unfold the CSS.
Prepare the CSS, two (2)
small sterile bags, and
three (3) pieces of tape on
a flat surface in the sterile
field.
3.
Position the HFC on flat work surface in the sterile field.