Use Instructions
PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 70 of 101
Table 35 Imaging Mode User Interface Icons
# Icon Name Description
10.
Ambient Light Sensor
Warning Icon
This icon appears overlaid in the top right
corner of the Imaging Field of View when the
Ambient Lighter Warning has been accepted
11.
Range Finder (RF)
These range icons appear in the bottom center
of the Imaging Field of View when the camera
sensor is too close or too far from the imaging
target.
12.
Temperature Gauge
The temperature gauge communicates relative
temperature to the operator.
6.2.2.2.1. Temperature Gauge
The temperature gauge communicates relative temperature to the operator. When the gauge approaches
full, the HFC is approaching a maximum operating temperature of 41°C. When operating temperature of
41°C is reached, the HFC automatically disables white light or blue-violet LEDs.
6.2.2.2.2. Range Finder (RF)
During White Light Imaging or Fluorescence Imaging, the Range Finder will provide real time instruction
to “Move back” from or “Move closer” to the imaging target. When the optimal imaging range has been
met, the Range Finder is not present on the display screen.
Capturing images or videos when the Range Finder is present may result in reduced image quality.
6.2.2.2.3. Circle of Interest
Align imaging targets during White Light Imaging and Fluorescence Imaging within the Circle of Interest.
Imaging within the Circle of Interest ensures the imaging target is uniformly illuminated during White Light
Imaging and Fluorescence Imaging.
Do not use fluorescence information observed outside the Circle of Interest when interpreting
fluorescence images per Section 6.2.2.2.5.2.
6.2.2.2.4. White Light Imaging
Accessed through Imaging Mode, White Light Imaging can only be performed after creating a new
patient session (Section 6.2.2.1.1), continuing a previous session (Section 6.2.2.1.2), or beginning a
Demo Session (Section 6.2.2.1.3).