Use Instructions
PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 54 of 101
Table 18 Handheld Fluorescence Camera Sterilization Instructions
9.
Completely dry the HFC by wiping with a sterile lint-free cloth.
10.
Visually inspect the device to verify there are no signs of corrosion, discoloration, pitting, or cracked
of seals. Visually inspect the HFC to ensure the optical glass (Item X in Figure X) is free of any visible
material (e.g. fibers) or smudges (e.g. water stains). Use sterile lint-free cloth to wipe the optical
glass if required.
11.
In the sterile area of the dedicated workspace, wrap the HFC in a sterile blue cloth.
12.
The HFC is now sterile and ready for use.
5.2. ECH – External Communications Hub Cleaning and Disinfection
Warning
Do not clean or disinfect the ECH or any of its components when energized. Unplug
all cable connections before cleaning or disinfecting. Cleaning or disinfecting when
energized may be hazardous to the operator and/or destructive to the device(s).
Caution
Prior to disinfection, ensure surfaces are cleaned to remove organic matter as
described in the sections below.
Thoroughly clean and disinfect all External Communications Hub component surfaces including the ECH
Display, ECH PC, ECH Pad, ECH Stand, ECH Power Supplies, and cables using Intermediate Level
Disinfection Wipes.