5.1. HFC Handheld Fluorescence Camera Cleaning and Sterilization
Before use, the Handheld Fluorescence Camera must be sterilized.
SBI ALApharma Canada Inc. has validated and therefore recommends the use of mild soap and water as
an effective cleaning agent before sterilizing the Handheld Fluorescence Camera with glutaraldehyde
(2.5% solution).
Following immediate use and before re-sterilization, SBI ALApharma Canada Inc. recommends to clean
and disinfect the Handheld Fluorescence Camera with Intermediate Level Disinfection wipes.
5.1.1. Cleaning the HFC
Table 17Handheld Fluorescence Camera Cleaning Instructions
1.
Prior to sterilization, the HFC must be thoroughly cleaned with mild soap and water to remove
organic matter.
2.
Wear appropriate PPE (e.g. impervious gloves such as latex
gloves, masks, hair bonnets, gowns).
3.
Before cleaning, ensure that the hatch covering the USB-c port
is closed and secured in place.