Use Instructions

PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 49 of 101
cross-contamination, and patient or user exposure to microbiological contamination as described in
Section 5.1.
The External Communications Hub described in sections 2.2.2 and 6.3.1.2 positioned in the sterile field is
cleaned and disinfected using intermediate level disinfection wipes as described in section 5.2.
The Protective Storage Cradle described in sections 0 and 6.3.5 positioned in the sterile field is cleaned
using intermediate level disinfection, and steam sterilized as described in section 5.5.
The Contact Charging Device described in sections 2.2.1 and 6.3.5 positioned in the non-sterile field is
cleaned and disinfected using intermediate level disinfection wipes as described in section 5.3.
Table 16 Infrastructure, Materials, & Equipment Required for cleaning,
disinfection, and sterilization
#
Eagle V1.2
Imaging System
Component
Cleaning,
Disinfection, and
Sterilization
Method
Item Description
1.
Handheld
Fluorescence
Camera
Cleaning followed by
Chemical Sterilization
before use.
Clean and Disinfect
Before and After Use.
Dedicated workspace
divided into non-sterile
and sterile areas
To perform cleaning
and sterilization
Soap and Water To clean
Glutaraldehyde Chemical sterilant
Latex gloves, masks, hair
bonnets, gowns
Personal protective
equipment
Basin
To hold chemical
sterilant
Timer and Thermometer
To time and monitor
the sterilization
process
Lint-free cloth
To wipe down the
sterilized device
Sterile blue cloth
To wrap the sterilized
device