Specifications
Page | 9 516041-1118
ERROR MESSAGES
In order to ensure acceptable activation of DEFINITY®, VIALMIX®RFID is programmed to display error messages that alert
the user in the event that the activation cycle is outside the required duration or rate. The table below shows a
complete list of error
messages,
the associated VIALMIX®RFID
response,
the likely cause(s) of the
error,
and the action
required of the user.
Error
Associated VIALMIX®RFID Response
Likely Cause of
Message
Action Required by User
High Oscillation
Rate Error
High Oscillation Rate error message on
main screen (HE)
Audible alert
DEFINITY® vial has
shaken at higher
than acceptable
shake rate.
Do not use the DEFINITY®
vial. Vial was not acceptably
activated. Retain the vial and
call Lantheus Medical Imaging
(see page 15).
Press the STOP button to
clear the error.
Low Oscillation
Rate Error
Low Oscillation Rate error message on main
screen (LE)
Audible alert
DEFINITY® vial has
shaken at lower
than acceptable
shake rate.
Do not use the DEFINITY®
vial. Vial was not acceptably
activated. Retain the vial and
call Lantheus Medical Imaging
(see page 15).
Press the STOP button to
clear the error.
Motor Overcurrent
Motor Overcurrent error message on main
screen (MC)
Audible alert
Excess mechanical
loading or motor
failure.
Do not use the DEFINITY®
vial. Vial was not acceptably
activated. Retain the vial and
call Lantheus Medical Imaging
(see page 15).
Power cycle the unit using the
power switch on the back of
the unit to clear the error.
Thermal Fault
Thermal Fault error message on main
screen (OH)
Audible alert
Unit is overheating.
Do not use the DEFINITY®
vial. Vial was not acceptably
activated. Retain the vial and
call Lantheus Medical Imaging
(see page 15).
Power the unit off and allow
to cool. Once cooled, power
the unit on using the power
switch on the back of the unit.