Specifications
Page | 8 516041-1118
DEFINITY® is ready for use if VIALMIX®RFID shakes the vial uninterrupted for the full activation cycle with no error
messages displayed. The following screen will be displayed indicating a successful activation:
If the DEFINITY® vial does not shake for the full 45 second duration or at the acceptable rate, activation is not considered
successful and the vial must not be used. An error message will be displayed and an audible alarm will sound to alert
the user. Refer to page 9 for a list of error messages and the conditions they indicate.
Removing the DEFINITY® Vial
After successful activation of DEFINITY®, the vial can be removed for use. To remove the vial, open the cover and follow
these steps:
1. Press the lever down to unlatch the vial carrier then press the lever to the left to fully open the carrier.
2. Remove the DEFINITY® vial from the carrier.
3. Release the lever and the spring action will close the carrier.
4. Close the cover.
Powering Off VIALMIX®RFID
VIALMIX®RFID should be powered off at the end of the day. This is done by moving the switch on the back of the unit to
the "O" position.
USAGE COUNTER
VIALMIX®RFID is equipped with an internal usage counter which counts the total number of activation cycles performed.
A Usage Count Alert will be displayed on the VIALMIX®RFID main display (see page 11) when > 95% of the programmed
limit of activation cycles has been reached. Please contact Lantheus Medical Imaging Customer Service IMMEDIATELY
upon receiving this alert to request a new VIALMIX®RFID. Once the usage counter reaches 100% of the programmed
limit of activation cycles, a Usage Count Error message will be displayed on the VIALMIX®RFID main display (see page
11) and VIALMIX®RFID will cease to operate.