Specifications
Page | 3 516041-1118
SAFETY PRECAUTIONS
IMPORTANT: Read this section carefully before plugging in and operating the unit. VIALMIX®RFID
must be used in strict accordance with these instructions. Lantheus Medical Imaging has no liability
for any damage arising from any other use of this unit.
WARNING: Do not operate VIALMIX®RFID and RFID-tagged vials within 6 inches (15cm) of a pacemaker
and/or defibrillator.
WARNING: Use of VIALMIX
®
RFID adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the VIALMIX
®
RFID, including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could
result.
VIALMIX®RFID must only be operated with the lid closed and intact.
Do not store vials under the lid. Only the DEFINITY® vial placed in the cradle intended for immediate activation
should be in the enclosure.
VIALMIX®RFID must be placed on a level surface and away from the edge of the surface.
A dangerous condition can result from condensation build-up. Before using VIALMIX®RFID, allow the unit to
reach room temperature, particularly when moving it from a colder to a warmer environment.
To safely operate VIALMIX®RFID, insert the main plug only into an appropriate outlet with a grounded conductor.
VIALMIX®RFID can operate using 120 to 240 volts, 50 to 60 Hz power. If an extension cord is used, make sure that
the ground conductor is not broken.
In order to avoid electrical shock, DO NOT INSERT ANY OBJECTS INTO THE UNIT.
If, for any reason, it is possible that the safety of VIALMIX®RFID has been compromised, that unit must be
removed from operation and identified in such a way that it is not inadvertently operated again. For example, do
not attempt to use VIALMIX®RFID if it does not work as specified or is visibly damaged.
Keep solvents, inflammable liquids, and heat sources away from unit.
Do not allow cleansers to get into unit during cleaning; an electrical short or a dangerous malfunction may
occur. Do not immerse unit in any liquid.
If the unit is not operating as expected, please contact Lantheus Medical Imaging (see page 15). Do not open
unit housing or attempt to perform any maintenance or repair.
VIALMIX®RFID must be properly disposed of at the end of its service life. To dispose of VIALMIX®RFID, contact
Lantheus Medical Imaging (see page 15).