Specifications
Page | 10 516041-1118
Error
Associated VIALMIX®RFID Response
Likely Cause of
Message
Action Required by User
Motor Overvoltage
Motor Overvoltage error message on main
screen (MO)
Audible alert
Motor is operating
higher than standard
voltage condition
indicating potential
future failure.
Do not use the DEFINITY®
vial. Vial was not acceptably
activated. Retain the vial and
call Lantheus Medical Imaging
(see page 15).
Power cycle the unit using the
power switch on the back of
the unit to clear the error.
Motor Under
voltage
Motor Under voltage error message on main
screen (MU)
Audible alert
Motor is operating
lower than standard
voltage condition
indicating potential
future failure.
Do not use the DEFINITY®
vial. Vial was not acceptably
activated. Retain the vial and
call Lantheus Medical Imaging
(see page 15).
Power cycle the unit using the
power switch on the back of
the unit to clear the error.
RFID Read Error
RFID Read error message on main screen
(RE)
Audible alert
1. RFID Tag does
not have
correct string
of data.
2. RFID Tag is
unreadable.
3. No DEFINITY®
vial in the
cradle.
Re-position the DEFINITY® vial
in the carrier. If the error re-
occurs, remove the vial and
place a new DEFINITY® vial in
the carrier.
Do not use the DEFINITY®
vial. Retain the vial and call
Lantheus Medical Imaging
(see page 15).
Cover Open (before
start button
pressed or pressed
and doesn’t work)
Cover open icon displayed
Cover not closed
prior to start of
activation cycle.
Close the cover and press the
START button