Users Manual Part 1

61
Nerve Integrity Monitor
Troubleshooting
Should you encounter any diculty eliciting simulated responses from the NIM Vital system patient simulator, check the following:
Verify that the Stimulus Measured is approximately the same as the Stimulus Intensity.
Make sure the jumper cables are connected correctly between the SIMULATOR and PATIENT INTERFACE.
Adjust the EVENT THRESHOLD setting on the NIM Vital system.
Adjust the STIMULUS intensity on the NIM Vital system for adequate output.
Clean the stimulator contacts of debris.
Check the integrity of the stimulator or stimulus-dissection instrument and its connecting cable.
Check for a blown fuse in NIM Vital system patient interface and replace with a proper valued fuse (shown near fuse box).
Check for proper closure of fuse holder in the NIM Vital system patient interface.
Customer service information
For further information regarding the use of this product or to report any problems, please contact
Medtronic using the appropriate information provided on the blue and white contact
information card packaged with each device; or contact your local distributor.
Troubleshooting
Should you encounter any diculty eliciting simulated responses from the NIM Vital system patient simulator, check the following:
Verify that the Stimulus Measured is approximately the same as the Stimulus Intensity.
Make sure the jumper cables are connected correctly between the SIMULATOR and PATIENT INTERFACE.
Adjust the EVENT THRESHOLD setting on the NIM Vital system.
Adjust the STIMULUS intensity on the NIM Vital system for adequate output.
Clean the stimulator contacts of debris.
Check the integrity of the stimulator or stimulus-dissection instrument and its connecting cable.
Check for a blown fuse in NIM Vital system patient interface and replace with a proper valued fuse (shown near fuse box).
Check for proper closure of fuse holder in the NIM Vital system patient interface.
Customer service information
For further information regarding the use of this product or to report any problems, please contact
Medtronic using the appropriate information provided on the blue and white contact
information card packaged with each device; or contact your local distributor.
External display support
The NIM Vital system has the following external video port, but does not have any specic connection settings for a microscope
output:
HDMI
Display probe
Wireless display module
The wireless display module enables users to wirelessly replicate the NIM Vital console display onto any monitor with an HDMI input
port. Refer to the installation instructions packaged with the wireless display module.
NIM Vital Mute adapter (optional)
The NIM Vital mute adapter enables the NIM Vital console to be connected to an external muting detector.
1. Console Connector (below knobs) .
2. Connector to mute probe.
3. Tether.
NIM Muting detector (optional)
Refer to Precaution P7 for more information.
The optional Muting Detector Probe is designed to detect the presence of electronic noise from external devices (such as
electrocautery/electrosurgical unit) that may cause interference on the EMG monitor. The use of a mute probe could be considered if
there is excessive noise from the electrosurgical unit.
1. Anti-slide Ring.
2. Electronic Noise Detection Area.
3. Insulating Sleeve.
4. Cable Connector.
5. Ferrite.
1
2
3
1
2
3
4
5
FCC use only, not for Medical use