Users Manual Part 1

49
Nerve Integrity Monitor
Nervassure electrodes stimulator
1
2
3
4
5
6
1 Negative Electrode Jacks (electrode plug matches color
and wire is black)
2 Positive Electrode Jacks (electrodes, plug and wire are the
same color)
3 Paired Electrodes
4 Continuous Monitoring Electrodes Stimulator Output
(cathode)
5 STIM2/Nervassure Electrodes Stimulator Return (anode)
6 Electrode Ground
The patient interface box supports one Nervassure stimulator in the STIM 2 position. The Nervassure electrodes have a yellow
wire and black plug. Insert the black plug into the black (-) jack on the PI (STIM2). The Nervassure electodes deliver current to the
patient and act as the cathode.
The stimulus return electrode has a white wire and red plug. Insert the red plug into the white jack with a red ring (+). The stimulus
return electrode is required for stimulation and is referred to as the stimulus return or anode. Route the white (+) stimulus return
wire away from the channel 1- 4 electrode wires.
Patient Interface box and stimulator combinations
Note:
The NIM Vital STIM1 and STIM2/Nervassure are wired independent of each other with each requiring an output (stimulator) and
a return (red electrode with white wire).
By default STIM1 is on and STIM2 is o. STIM2 must be turned on manually (in Nervassure procedures the system turns on stim
2 when the user presses the baseline or adjust button). Refer to the “MONITORING mode or “SETUP mode topics.
By default STIM1 and 2 are Monopolar Probes and must be changed manually for Bipolar. Refer to the “Settings” topics.
Monopolar incrementing stimulator and Nervassure electrode stimulator
1
2
3
4
5
6
7
8
9
10
1 Stimulator Output, Nervassure Electrode (cathode)
2 STIM2/Nervassure Stimulator Return (anode)
3 Negative Electrode Jacks (electrode plug matches color
and wire is black)
4 Positive Electrode Jacks (electrodes, plug and wire are the
same color)
5 Paired Electrodes
6 Incrementing probe adapter
7 Remote Control Connector
8 Stimulator Output (cathode)
9 STIM 1 Stimulator Return (anode)
10 Electrode Ground
FCC use only, not for Medical use