Users Manual Part 1
42
Nerve Integrity Monitor
The following settings are available on the Nervassure EMG audio panel:
• Nervassure EMG audio - Mute the Nervassure EMG Audio using the selection option. If the user turns o the toggle button and
selects the Mute check box, then the system mutes the EMG audio for the sweeps initiated by a Nervassure stimulation pulse. If
the user turns on the toggle button (on is the default) and deselects the Mute check box, then the volume of the EMG audio for
Nervassure sweeps is controlled by the [+] and [-], or the slider bar.
• Voice - Nervassure values - if the user selects this toggle button check box, then the NIM periodically announces the last
Nervassure EMG value audibly. Use the drop-down menu to select the frequency of annunciation. Note: Voice – the user
cannot enable Stimulated EMG Values at the same time as Voice – Nervassure values. Therefore, when the user enables Voice –
Nervassure, the system automatically disables Voice – Stimulated EMG Values (found in Settings/Audio).
• Nervassure Stim 2 rates - you can adjust the following rates and baseline from the Nervassure stimulation panel using values
from their dropdown boxes:
— Slow rate
— Fast rate
— Baseline
• Trending Display - You can toggle on/o the following settings from the Nervassure display panel using the slide button:
— Show Nervassure Waveforms
— Trend Latency
— Show Entire Session
— Show as Mountain Graph
• View Color Conditions - You can view the Nervassure default alarm color conditions on this panel. The system marks any the
modied values with an asterisk.
Nervassure advanced settings
The Nervassure advanced settings enable you to adjust the following settings.
Note: If you adjust the advanced settings, you will change the default characteristics of your NIM Vital. Only someone with
knowledge of the advanced settings and how they impact your NIM Vital should make changes. If the user selects
, the system
displays a warning. The user must select [Cancel] or [Conrm] to return to settings or move forward to the advanced settings panel.
1
2
3
4
1 Advanced Settings button. Enables access to the Nervassure
advanced settings.
2 Nervassure Channels.
3 Nervassure Alert Conditions:
• Alert Type.
• Amplitude Threshold.
• Absolute Threshold.
• Latency Threshold.
• Monitoring study group adapted color scheme (2018, only in
Thyroid).
• Alarm on Best Channel (only in Thyroid).
4 Nervassure Alert Volumes:
• Yellow.
• Red.
The following advanced settings are available on the Nervassure advanced settings panel:
• Nervassure Channels - Select the EMG channels to trend.
• Nervassure Alert Conditions
— Alert type - Select one of the following alarm types from the dropdown menu:
— Amplitude only
— Latency only
— Amplitude AND latency
— Amplitude OR latency
— Amplitude Threshold - adjust the amplitude limit using or .
— Absolute Amplitude - adjust the absolute amplitude using the slider bar or .
— Latency threshold - adjust the latency limit using or .
• Nervassure Alert Volumes - adjust the red and yellow alert volumes using the slider bar or .