Users Manual Part 1
75
Nerve Integrity Monitor
Display default settings
NIM Vital
Procedure Name
Acoustic
Neuroma 4ch
Acoustic
Neuroma 2ch
Cochlear
Implant
Mastoid Thyroid
Thyroid with
Nervassure
Parotid
Neck
Dissection
Glomus
Lower
Extremity 4 Ch
Knee (Peroneal)
2Ch
Ankle (Tibial
Nerve) 2Ch
Event Threshold 100 µV
Event Capture Largest
Stimulus Rejection
Period
3.1mS 2.1 mS 3.1mS 5.0mS
Time Scale 50mS 25mS 50mS
Vertical Scale 500uV 1000 µV 500 µV 2000uV
Detect Artifacts Yes
Monitoring Audio EMG + Event Tones
Stimulus Delivery
Tone
Brief
STIM1 & 2 default settings
NIM Vital
Procedure Name
Acoustic Neuroma
4 Ch
Acoustic Neuroma
2 Ch
Cochlear
Implant
Mastoid Thyroid
Thyroid with
Nervassure™
Parotid Neck Dissection Glomus
Lower Extremity
4 CH
Knee (Peroneal)
2 CH
Ankle (Tibial
Nerve) 2 CH
STIM1 Name STIM1
STIM1 Mode Regular
STIM1 Pulse
Width
100uS
STIM1 Current 0.8mA 1.0mA 0.8 mA 1.0mA 0.8mA 1.0mA
STIM1 Warning
Level
3.0 mA 12.0mA
STIM1 Rate 4/sec
STIM2 Name STIM2
Vagus STIM
Nervassure
STIM2
STIM2 Mode Regular
STIM2 ON/OFF OFF ON OFF
STIM2 Pulse
Width
100uS
STIM2 Current 0.0mA 1.0 mA 0.0mA
STIM2 Warning
Level
3.0mA Variable 3.0mA 12.0mA
STIM2 Rate 4/sec 1/sec 4/sec
FCC use only, not for Medical use