Users Manual Part 1

64
Nerve Integrity Monitor
Additional mute probe adapter information
To remove the mute probe adapter from the console, squeeze the two tabs on the side of the adapter to release it.
To tether the mute probe adapter to the muting detector, do the following:
1. Loop the mute probe adapter tether around the muting detector.
2. Take the end of the mute probe adapter marked with “Medtronic” and slip it through the tether, while going around the muting
detectors cable.
Muting
The Muting Detector is normally clamped around the active cable of an electrosurgical instrument that would otherwise interfere
with monitoring. Do not include the grounding pad cable in the jaws of the Muting Detector. It may also be connected to other
equipment cables, such as electrocautery, ultrasonic debulking devices (for example, CUSA), or other external devices, which
generate interfering signals. Always attach the Muting Detector close to the output jacks of the electrosurgical or other unit rather
than near the hand piece the surgeon uses. More than one output cable may be clamped within the Muting Detector jaws at one
time.
Monopolar electrosurgical instrument clamping
Power level for satisfactory muting performance of electrosurgical unit:
Monopolar Electrosurgical Unit
5-100W
STIM Bur Guard
The STIM Bur Guard provides stimulating current to standard Medtronic burs in both static and
dynamic modes. The STIM Bur Guard locks onto any Visao handpiece, while the cable connects to
a port on the IPC System or XPS 3000 (REF 1897102BF). A separate cable then connects to the
NIM Patient Interface box. Wires within the STIM Bur Guard make contact with the uncoated bur
and carry the stimulating current to the bur tip.
Prior to using this device, you must be familiar with the NIM Systems Users Guide, the Bur
manual, the STIM Bur Guard Users Guide, and the IPC System or XPS 3000 (REF 1897102BF) User’s
Guide.
Compatible equipment:
XPS 3000 Console (1897102BF)
IPC Console
For more information related to the STIM Bur Guard system contact your local Medtronic representative.
Accessories / parts list
System components & accessories
(See Medtronic Catalog for product details Note: online some numbers require a dash xx-xxxxx)
P/N Description P/N Description
NIM4CM01 NIM Vital Console NIM4CC01 NIM Vital Cart
NIM4CS01 NIM Vital Storage Case NIM4CB01 NIM Vital Battery Console
Mainframe accessories
NIM4CPB1 NIM Vital Patient Interface (4 channel) NIM4CAV1 NIM Vital Video/Audio Recording Adaptor Kit
NIM4CFU1 Fuse Kit, NIM4CFU1 NIM Vital Console NIM4CC02 NIM Vital Power Isolator for Printer
8253075 FUSE KIT 8253075 NIM 3.0 INTFC 32MA NIM4CPB2 NIM Vital Patient Interface Connector Cable
NIM4CC03 International NIM Vital Power Isolator for Printer 8220325 Muting Probe NIM
NIM4CD01 NIM Vital Wireless Display Module NIMCAD400 NIM Incrementing Probe Adapter
8253030 Keyboard NIM4CAM1 INTFC Adaptor Mute Probe
EMG Endotracheal Tubes
8229506 NIM EMG Endotracheal Tube, 6mm 8229709 NIM Trivantage EMG Endotracheal 9.0 MM ID
8229507 NIM EMG Endotracheal Tube, 7mm 8229735 3PK NIM Trivantage 5 MM
8229508 NIM EMG Endotracheal Tube, 8mm 8229736 3PK NIM Trivantage 6 MM
8229705 NIM Trivantage EMG Endotracheal 5.0 MM ID 8229737 3PK NIM Trivantage 7 MM
8229706 NIM Trivantage EMG Endotracheal 6.0 MM ID 8229738 3PK NIM Trivantage 8 MM
8229707 NIM Trivantage EMG Endotracheal 7.0 MM ID 8229739 3PK NIM Trivantage 9 MM
8229708 NIM Trivantage EMG Endotracheal 8.0 MM ID
Monopolar Stimulating probes
8225101 Standard Prass Flush-Tip Stimulator Probe, 5/Box 8225278 Ball-Tip Monopolar Stim Probe 1mm (tips only), 10/Box
8225110 Standard Prass Flush-Tip Stim Probe (tips only), 10/Box 8225251 Ball-Tip Monopolar Stim Probe 2.3mm (tips only), 10/Box
8225275 Ball-Tip Monopolar Stimulator Probe 1mm, 1/Box 8225277 Yingling Flex-Tip Monopolar Stimulating Probe, 5/Box
8225825 Incrementing Probe with Standard Prass Tip, 3/Box 8225051 Monopolar Stimulator Handles, 10/Box
8225276 Ball-Tip Monopolar Stimulator Probe 2.3mm, 1/Box 8225490 Incrementing Probe, Ball Tip 1mm 3/Box
FCC use only, not for Medical use