Users Manual Part 1
47
Nerve Integrity Monitor
After conrming the electrodes are properly placed in the patient and connected to the patient interface box, perform an Electrode
Check on the NIM Vital system. This may determine if the electrodes need to be re-inserted for better electrical contact to help
optimize monitoring.
If you are using the wired patient interface box, lay the patient interface box cable out of busy trac patterns and secure it to the
oor as needed.
Stimulators
The NIM Vital system Stimulator output supports the use of both monopolar and bipolar stimulating probes with a range of 0-50
mA.
Monopolar Incrementing probe
The single use incrementing probe provides the surgeon with the means to adjust the stimulation current and print or save the
monitoring screen shots from the sterile eld as well as other features.
The patient interface supports one incrementing probe. The smaller plug is color-coded black and plugs into the black (-) jack on the
patient interface box. The larger plug is also color-coded black and plugs into the large black (STIM CONTROL) jack on the patient
interface box. If you are using the incrementing probe adapter, plug the incrementing probe into the adapter and then plug the
adapter into the patient interface box. The monopolar stimulating probe delivers current to the patient and acts as the cathode. The
red electrode with white wire (+) is required for monopolar stimulation. It is referred to as the stimulus return or anode. Plug the red
connector of the red electrode with white wire into the white jack with a red ring (+). Route the white (+) stimulus return wire away
from the channel electrode wires.
1
2
3
4
5
6
7
8
9
1 Monopolar Probe with Remote Stimulus Adjust
(Incrementing Probe) Handle
2 Remote Control Connector
3 Stimulator Output (cathode)
4 Negative Electrode Jacks (electrode plug matches color and
wire is black)
5 Positive Electrode Jacks (electrodes, plug and wire are the
same color)
6 Electrode Ground
7 STIM1 Stimulator Return (anode)
8 Paired Electrodes
9 Incrementing probe adapter
Note: The Incrementing probe connector (2) has three indentations in the connector sleeve and one-o center key that must be
aligned with the incrementing probe adapter (3) or the connector may be damaged. Align white arrow (1) on side of Incrementing
probe connector with the black arrow on incrementing probe adapter.
2 3
1