Users Manual Part 1
40
Nerve Integrity Monitor
If the Nervassure values go above or below the set limits, the stimulation point changes to yellow or red depending on the limits the
value crosses. The yellow and red zones have a unique tone associated with their state. Refer to the “Alarms and alarm criteria” t opic
for additional information.
Thyroid with Nervassure specic features
The thyroid with Nervassure procedure has Best Channel mode turned on by default. Refer to the “Nervassure advanced settings”
topic for additional information. Uncheck the Thyroid with Nervassure mode to view.
Muting a Nervassure alarm
When a plotted point falls within the set alarm band, the system displays an alarm button on screen. You can mute this Nervassure
alarm on the Monitoring screen using the alarm button. There are three levels of alarm with their availability determined by
procedure settings – a red, yellow, and loss of signal. Each level has a separate alarm button with mute functionality that mutes only
that level (so muting yellow and the trend moves to the loss of signal limit still sounds the red alarm until that one is muted.
1. Press .
The system mutes the alarm and displays a muted alarm.
2. To unmute the alarm, press the muted alarm button.